Can simultaneously measure parameters such as frequency stability, aging, phase drift, etc
Measurement frequency range: 0.5MHz~30MHz; 100MHz±10MHz
Frequency Stability:<2E-14 (typical value: 1.2E-14)
Multi channel: Standard configuration includes 4 channels, expandable to 16 channels(including reference input)
The measured signal and the reference signal do not need to be on the same frequency
Sampling time: 1ms~1000s
Equipped with an upper computer, the measurement results are displayed and stored in real-time
Product Features | Specification name | Performance Parameters | Indicator parameters |
Measuring indicators | Reference input | 0.5MHz~30MHz 90MHz~110MHz | Signal impedance: 50 Ω Signal waveform: Both sine wave and square wave are supported |
Number of Reference Channels | 1st Channel | ||
Measured frequency | 0.5MHz~30MHz 90MHz~110MHz | ||
Number of tested paths | 4-Channel/8-Channel/16-Channel | ||
Stability | Regular model: <5E-14@1s | ||
Professional model: <2E-14@1s | |||
Phase comparison accuracy | <100fs (gate time 1s) | ||
Sampling time | 1ms, 10ms, 100ms, 1s, 10s, 100s, 1000s | ||
Power measurement resolution | 0.1dB | ||
Measurement configuration | Automatic input frequency recognition | ||
Supply voltage | Supply voltage | 220VAC 50Hz | |
Power consumption | Chassis: ≤100.0W | ||
Power interface | International power plugs | ||
Ambient temperature | Operating temperature | -20ºC ~ +50ºC | |
Storage temperature | -40ºC ~ +85ºC | ||
Weight and Size | Weight | <15kg (whole machine) | |
Size | 1U/4 channels, 2U/8 channels, 2U/16 channels |